FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3862426 · Received June 10, 2014

Report

Report Number
2649622-2014-06928
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
March 21, 2014
Report Date
March 21, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY- THE FULL LEAD WAS RETURNED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND.

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE ATRIAL LEAD WOULD NOT ALLOW THE STYLET TO BE INSERTED. A DIFFERENT LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338160 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 407652

Patients

Seq Age Sex Outcome Treatment
1 00083 YR