SIGMA 300 DR
Report
- Report Number
- 3004209178-2014-11076
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- April 14, 2014
- Report Date
- April 14, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NVZ
- PMA / PMN Number
- P980035
- Removal / Correction Number
- Z-1516-2009
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
Narratives
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 5024M-58 LEAD, IMPLANTED: (B)(6) 1995. (B)(4).
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. PRELIMINARY ANALYSIS REVEALED NO OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES.
IT WAS REPORTED THAT THE PATIENT HAD DIZZINESS AND BRADYCARDIA DUE TO THE DEVICE HAVING A PACING AND SENSING ISSUE. IT WAS NOTED THAT THE DEVICE HAD NO CAPTURE AND NO SENSING. THE DEVICE WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 338540 | SIGMA 300 DR | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | SDR303B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00059 YR | Hospitalization| L| R | 5524M-53 LEAD |