FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3862368 · Received June 10, 2014

Report

Report Number
2649622-2014-06958
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 8, 2014
Report Date
April 8, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. THE DISTAL CONDUCTOR OF THE LEAD WAS EXTRINSICALLY OVER-ROTATED, AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THE VENTRICULAR LEAD WOULD NOT HOLD THE INITIAL POSITION, AND AFTER MULTIPLE ATTEMPTS, THE HELIX WOULD NO LONGER RETRACT. THE LEAD WAS NOT USED. A SECOND LEAD WAS ATTEMPTED, AND ALSO WOULD NOT HOLD THE INITIAL POSITION IN MULTIPLE PLACEMENTS. THE LEAD WAS EVENTUALLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
337827 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI58

Patients

Seq Age Sex Outcome Treatment
1 00058 YR