FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY

MDR report key: 3862336 · Received June 10, 2014

Report

Report Number
3008973940-2014-00208
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 8, 2014
Report Date
April 8, 2014
Manufacturer
MEDTRONIC SINGAPORE OPERATIONS
Product Code
OJX
PMA / PMN Number
P080006
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE U.S WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. CONCOMITANT PRODUCTS: 1888TC LEAD IMPLANTED (B)(6) 2010; 7122 LEAD IMPLANTED (B)(6) 2010.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE STYLET BECAME LODGED IN THE NEWLY IMPLANTED LEFT VENTRICULAR (LV) LEAD AFTER CUTTING THE DELIVERY SYSTEM AND REQUIRED GREAT FORCE TO REMOVE. THE STYLET TOOK ON A SINUSOIDAL SHAPE WHEN REMOVED. THE PHYSICIAN BELIEVED THAT THIS LEFT PENT UP TORQUE IN THE LEAD AND MAY HAVE CONTRIBUTED TO THE LEAD DISLODGING BY THE FOLLOWING MORNING. THE LEAD WAS SUBSEQUENTLY REMOVED. A NEW LEAD WAS SUCCESSFULLY PLACED IN THE TARGET POSITION. THE STYLET BECAME STUCK FOLLOWING SLITTING AND REMOVAL OF THE DELIVERY SYSTEM. SIGNIFICANT FORCE WAS REQUIRED TO SLOWLY PULL THE STYLET OUT OF THE LEAD AND A CONSTANT FRICTION-TYPE NOISE COULD BE HEARD, SIMILAR TO THAT OF UNROLLING MASKING TAPE. ONCE REMOVED, THRESHOLDS REMAINED UNCHANGED AND THE LEAD REMAINED IN THE TARGET POSITION. HOWEVER, LEAD SLACK IN A FIGURE EIGHT SHAPE WAS OBSERVED IN THE CORONARY SINUS BODY, PRESUMABLY FROM THE RELEASE OF TORQUE FROM THE PROLONGED TENSION DURING STYLET REMOVAL. CONCERN WAS THEN EXPRESSED AS TO WHETHER THE STYLET WAS OF THE CORRECT DIAMETER FOR THE LEAD LUMEN. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338419 ATTAIN ABILITY DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MEDTRONIC SINGAPORE OPERATIONS 419688

Patients

Seq Age Sex Outcome Treatment
1 00073 YR DTBA201 ICD