FDA Adverse Event Injury Summary report: N

CONSULTA CRT-D

MDR report key: 3862306 · Received June 10, 2014

Report

Report Number
3004209178-2014-11092
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 2, 2014
Report Date
April 2, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Removal / Correction Number
Z-0111-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE . IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 694758, LEAD, IMPLANTED: (B)(6) 2010; 5076-52, LEAD, IMPLANTED: (B)(6) 2010. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 91% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD REACHED BATTERY DEPLETION EARLY. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
337743 CONSULTA CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D224TRK

Patients

Seq Age Sex Outcome Treatment
1 00080 YR Hospitalization| R 419688 LEAD