FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3862302 · Received June 10, 2014

Report

Report Number
2649622-2014-06983
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 15, 2014
Report Date
January 29, 2015
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. ANALYSIS INDICATED THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. THE SVC (SUPERIOR VENA CAVA) DEFIBRILLATION COIL DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. THE PROXIMAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN A FIELD ACTION. BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT DETERMINE THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT HIGH IMPEDANCE LEVELS WERE DISCOVERED ON A RIGHT VENTRICULAR (RV) LEAD. A POTENTIAL LEAD FRACTURE IS SUSPECTED. THE LEAD CURRENTLY REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

THE LEAD HAS NOW BEEN REMOVED AND REPLACED WITH A NEW LEAD DUE TO PREVIOUSLY REPORTED FRACTURE. NO PATIENT COMPLICATIONS HAVE BEEN R EPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338384 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00058 YR Hospitalization| R D274VRC, IMPLANTABLE CARDIOVERTER DEFIBRILLATOR