FDA Adverse Event
Malfunction
Summary report: N
CARELINK
MDR report key: 3862202
·
Received June 10, 2014
Report
- Report Number
- 2182208-2014-01708
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- March 17, 2014
- Report Date
- April 14, 2014
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- KRG
- PMA / PMN Number
- P890003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS FOUND THAT THE DEVICE WOULD NOT INTERROGATE AND THE UPLINK FUNCTIONAL TESTS WERE OUT OF SPECIFICATION DUE TO THE CABLE BEING OUT OF ELECTRICAL SPECIFICATION. IT WAS ALSO NOTED THAT THE LABEL WAS MISSING THE LAMINATE BACKING. (B)(4).
Description of Event or Problem · 1
THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 342692 | CARELINK | PROGRAMMER, PACEMAKER | KRG | MEDTRONIC, INC. | 2067 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |