FDA Adverse Event Malfunction Summary report: N

CAPSURE

MDR report key: 3862199 · Received June 10, 2014

Report

Report Number
2649622-2014-07013
Event Type
Malfunction
Date Received
June 10, 2014
Report Date
April 18, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P830061
Removal / Correction Number
N-029/030-4
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND THE PROXIMAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION. (B)(4).

Description of Event or Problem · 1

THE RIGHT VENTRICULAR (RV) LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338026 CAPSURE ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 4004M52

Patients

Seq Age Sex Outcome Treatment
1 00060 YR 8360 IPG