CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2014-07032
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- April 30, 2014
- Report Date
- April 30, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: A 4076 LEAD, IMPLANTED: (B)(6) 2014. (B)(4).
PRODUCT EVENT SUMMARY - THE FULL LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. HOWEVER, THE PROXIMAL CONDUCTOR OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE.
IT WAS REPORTED THAT THE ATRIAL LEAD DISLODGED THE DAY AFTER IMPLANT. THE PHYSICIAN WAS UNABLE TO REVISE THE LEAD AND IT WAS EXPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 341546 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00083 YR | Hospitalization| R | A2DR01 IPG |