FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3862156 · Received June 10, 2014

Report

Report Number
2649622-2014-07032
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 30, 2014
Report Date
April 30, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: A 4076 LEAD, IMPLANTED: (B)(6) 2014. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY - THE FULL LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. HOWEVER, THE PROXIMAL CONDUCTOR OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD DISLODGED THE DAY AFTER IMPLANT. THE PHYSICIAN WAS UNABLE TO REVISE THE LEAD AND IT WAS EXPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341546 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00083 YR Hospitalization| R A2DR01 IPG