FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY PLUS

MDR report key: 3862107 · Received June 10, 2014

Report

Report Number
2649622-2014-07053
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 7, 2014
Report Date
April 7, 2014
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCTS: 5076-52 LEAD IMPLANTED: (B)(6) 2010. (B)(4).

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND; FULL LEAD RETURNED AND ANALYZED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT ATTEMPT, THE LEAD WAS UNSTABLE DUE TO THE SIZE OF THE VESSEL. HIGH THRESHOLDS WERE ALSO NOTED, AND THE LEAD WAS REPORTED TO CAUSE DIAPHRAGMATIC STIMULATION. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342429 ATTAIN ABILITY PLUS DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 429688

Patients

Seq Age Sex Outcome Treatment
1 00080 YR 694765 LEAD