FDA Adverse Event Injury Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3862046 · Received June 10, 2014

Report

Report Number
2649622-2014-07065
Event Type
Injury
Date Received
June 10, 2014
Date of Event
March 5, 2014
Report Date
March 5, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY WAS PERFORMED AND NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT IMMEDIATELY FOLLOWING IMPLANTATION, AFTER THE POCKET HAD BEEN CLOSED, THE RIGHT VENTRICULAR (RV) LEAD EXPERIENCED A GRADUAL RISE IN BOTH IMPEDANCE AND CAPTURE THRESHOLDS. FLUOROSCOPY PICTURES WERE TAKEN TO EVALUATE THE RV LEAD AND THE HELIX OF THE LEAD APPEARED TO NOT BE FULLY DEPLOYED. THE POCKET WAS REOPENED AND THE RV LEAD WAS REPOSITIONED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342217 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI58

Patients

Seq Age Sex Outcome Treatment
1 00079 YR Hospitalization| R 5086MRI52 LEAD