FDA Adverse Event Injury Summary report: N

ADAPTA DR

MDR report key: 3861940 · Received June 10, 2014

Report

Report Number
9614453-2014-01411
Event Type
Injury
Date Received
June 10, 2014
Date of Event
June 26, 2013
Report Date
March 31, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NVZ
PMA / PMN Number
P980035
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXPERIENCED "A LACK OF STIMULATION." IT WAS FURTHER REPORTED THAT THE LEAD EXHIBITED HIGH IMPEDANCES AND HIGH THRESHOLDS. THE LEAD AND DEVICE WERE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
337296 ADAPTA DR PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ IPG MFG SWITZERLAND ADDRL1

Patients

Seq Age Sex Outcome Treatment
1 00076 YR Hospitalization| R 5076-52