FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3861813 · Received June 10, 2014

Report

Report Number
2649622-2014-07174
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 28, 2014
Report Date
April 28, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE PARTIAL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED AND NO ANOMALIES WERE FOUND. THE PROXIMAL AND DISTAL CONDUCTORS OF THE LEAD BECAME EXTRINSICALLY FRACTURED DUE TO A CUT. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE. THE LEAD WAS DESTRUCTIVELY ANALYZED. NO ANOMALIES COULD BE ATTRIBUTED TO THE COMPLAINT AT THIS TIME.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: SEDR01, IPG, IMPLANTED: (B)(6) 2007. (B)(4).

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT A CLINICAL DATA REVIEW REVEALED THAT THE RV LEAD IMPEDANCE TREND SHOWS A SUDDEN INCREASE IN IMPEDANCE OVER A PERIOD OF A FEW WEEKS TO OUT OF RANGE BETWEEN (B)(6) 2014. THE THRESHOLD TREND ALSO SHOWS A SUDDEN INCREASE IN THRESHOLD TO OVER 2.5 VOLTS OVER THE SAME TIME PERIOD FROM A STABLE THRESHOLD OF 0.625 VOLTS. THE VHR EPISODE LOG IS FULL OF EPISODES (FROM PRIOR TO LAST SESSION W/O EGM) FROM (B)(6) WITH THE FASTEST V RATE BEING OVER 450 BPM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS FRACTURED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340334 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-58

Patients

Seq Age Sex Outcome Treatment
1 00073 YR Hospitalization| R 1888TC ST JUDE LEAD