FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3861754 · Received June 10, 2014

Report

Report Number
2649622-2014-07190
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 8, 2014
Report Date
April 8, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A CUT, AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT; FULL LEAD RETURNED AND ANALYZED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO USE, THE ANCHORING SLEEVE WAS MOVED AND SEEMED TO BE STUCK. WHEN IT FINALLY RELEASED, THE PHYSICIAN OBSERVED A BREAK IN THE LEAD INSULATION. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341902 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00066 YR