FDA Adverse Event Malfunction Summary report: N

MAXIMO II CRT-D

MDR report key: 3861731 · Received June 10, 2014

Report

Report Number
3004209178-2014-11156
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 2, 2014
Report Date
April 2, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Removal / Correction Number
Z-0112-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION. BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT DETERMINE THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION. CONCOMITANT PRODUCTS: 1581 ST. JUDE LEAD, IMPLANTED: (B)(6) 2005; 1688TC ST. JUDE LEAD, IMPLANTED: (B)(6) 2005. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 88% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MEASURED BATTERY VOLTAGE WAS BELOW THE VOLTAGE NEEDED TO TRIP THE RECOMMENDED REPLACEMENT INDICATOR; HOWEVER, NO INDICATOR HAD BEEN TRIGGERED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

THE DEVICE WAS EXPLANTED AND REPLACED AFTER ELECTIVE REPLACEMENT INDICATOR WAS TRIGGERED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340059 MAXIMO II CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D284TRK

Patients

Seq Age Sex Outcome Treatment
1 00055 YR Hospitalization| R 4196-78 LEAD