FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 3861711
·
Received June 10, 2014
Report
- Report Number
- 2649622-2014-07194
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- April 17, 2014
- Report Date
- April 17, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: ADDR01, IPG, IMPLANTED: (B)(6) 2011. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS, OVERSENSING, AND IMPEDANCE LEVELS OF GREATER THAN 9,999 OHMS RESULTING FROM A CONFIRMED FRACTURE. THE RA LEAD WAS ABANDONED AND REPLACED SUCCESSFULLY WITH A PLAN TO CAP AT A LATER DATE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 342066 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00081 YR | Hospitalization| R | 507658 LEAD |