FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3861685 · Received January 13, 2014

Report

Report Number
1314492-2014-02623
Event Type
Malfunction
Date Received
January 13, 2014
Date of Event
November 1, 2013
Report Date
December 17, 2013
Manufacturer
BAXTER HEALTHCARE CORP.
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURER REF NO.: (B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF COMPLAINT (B)(4). THE IDENTIFIED "DOOR JAMMED" ALARMS WERE CONFIRMED THROUGH THE EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADDITIONAL INFO IS RECEIVED, A FOLLOW UP WILL BE SENT. THE LOWER AUXILIARY FLEX ASSEMBLY WAS REPLACED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED SPECTRUM INFUSION PUMP, 2 OCCURRENCES OF "DOOR JAMMED" ALARM WERE IDENTIFIED IN THE EVENT HISTORY LOG. ANY PT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION IS UNKNOWN SINCE THESE ITEMS WERE FOUND DURING EVALUATION OF THE DEVICE. NO ADDITIONAL INFO IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
29038 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP. 35700

Patients

Seq Age Sex Outcome Treatment
1