MAXIMO II DR
Report
- Report Number
- 3004209178-2014-11165
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- March 28, 2014
- Report Date
- March 28, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- LWS
- PMA / PMN Number
- P980016
- Removal / Correction Number
- Z-0113-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 87% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL. THIS DEVICE WAS INCLUDED IN A FIELD ACTION AND RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT : 6947-65, LEAD, IMPLANTED (B)(6) 2009. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE DEVICE BATTERY VOLTAGE DROPPED OVER A THREE DAY PERIOD, THEN RECOVERED AND DECREASED AGAIN. PERFORMANCE DATA FROM THE DEVICE WAS ANALYZED AND THE SUDDEN BATTERY DEPLETION MAY BE ATTRIBUTED TO CONTINUED WIRELESS TELEMETRY, WHICH WAS ACTIVE FOR ABOUT 7 DAYS. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 341503 | MAXIMO II DR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D284DRG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00071 YR | Hospitalization| R | 4076-52 LEAD |