ATTAIN ABILITY PLUS
Report
- Report Number
- 2649622-2014-07216
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- April 9, 2014
- Report Date
- April 9, 2014
- Manufacturer
- MPRI
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND THE OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO PULLING/STRETCHING/OVERSTRESS.
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE OUTER INSULATION DISCONNECTED FROM THE DISTAL END OF THE LEAD PIN INSULATION EXPOSING INNER CONDUCTORS. IT WAS NOTED THAT NO PRESSURE/PULLING WAS APPLIED TO THE LEAD TO CAUSE THIS ISSUE. A NEW LEAD WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 339691 | ATTAIN ABILITY PLUS | DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE | OJX | MPRI | 429688 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |