FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY PLUS

MDR report key: 3861637 · Received June 10, 2014

Report

Report Number
2649622-2014-07216
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 9, 2014
Report Date
April 9, 2014
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND THE OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO PULLING/STRETCHING/OVERSTRESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE OUTER INSULATION DISCONNECTED FROM THE DISTAL END OF THE LEAD PIN INSULATION EXPOSING INNER CONDUCTORS. IT WAS NOTED THAT NO PRESSURE/PULLING WAS APPLIED TO THE LEAD TO CAUSE THIS ISSUE. A NEW LEAD WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
339691 ATTAIN ABILITY PLUS DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 429688

Patients

Seq Age Sex Outcome Treatment
1