FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3861587 · Received June 10, 2014

Report

Report Number
2649622-2014-07228
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 1, 2014
Report Date
April 1, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 4194-88, LEAD, IMPLANTED (B)(6) 2010; D224TRK, ICD, IMPLANTED (B)(6) 2010. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT RECEIVED INAPPROPRIATE SHOCKS DUE TO A RIGHT VENTRICULAR LEAD FRACTURE. THE LEAD WAS OVERSENSING NOISE AND A LEAD INTEGRITY ALERT WAS TRIGGERED. THE LEAD WAS CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342861 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00065 YR Hospitalization| L| R 4076-52 LEAD