FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 3861505 · Received June 10, 2014

Report

Report Number
2182208-2014-01762
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 2, 2014
Report Date
April 29, 2014
Manufacturer
MEDTRONIC, INC.
Product Code
KRG
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT, THE SYSTEM FAN WAS REPLACED. IT WAS ALSO NOTED THAT THE PROGRAMMER FAILED THE RIGHT ANTENNA TEST, AND THE ELECTROCARDIOGRAM (ECG) CONNECTOR WAS LOOSE ON THE LINK ELECTRONIC MODULE (LEM) CIRCUIT BOARD. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER HAD A NOISY SYSTEM FAN. THE PROGRAMMER WAS RETURNED FOR SERVICE. IT WAS ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
343012 CARELINK PROGRAMMER, PACEMAKER KRG MEDTRONIC, INC. 2090

Patients

Seq Age Sex Outcome Treatment
1 229047 ANALYZER