FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 3861452
·
Received June 10, 2014
Report
- Report Number
- 2649622-2014-07275
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- January 13, 2014
- Report Date
- March 4, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND THE PROXIMAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. PRODUCTS: 5076-45 LEAD IMPLANTED: 2004 (B)(6). (B)(4).
Description of Event or Problem · 1
THE LEAD WAS RETURNED TO THE MANUFACTURER WITH NO INFORMATION. THE LEAD SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 343315 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-45 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00073 YR | Hospitalization| R | (B)(4) IPG |