Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND THE PROXIMAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. PRODUCTS: 5076-45 LEAD IMPLANTED: 2004 (B)(6). (B)(4).

Description of Event or Problem · 1

THE LEAD WAS RETURNED TO THE MANUFACTURER WITH NO INFORMATION. THE LEAD SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
343315 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-45

Patients

Seq Age Sex Outcome Treatment
1 00073 YR Hospitalization| R (B)(4) IPG