FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3861419 · Received June 10, 2014

Report

Report Number
2649622-2014-07260
Event Type
Injury
Date Received
June 10, 2014
Report Date
March 21, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED WITH NO ANOMALIES FOUND. THE ANALYST NOTED THAT THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 5076-52 IMPLANTABLE PACING LEAD, IMPLANTED (B)(6) 2013. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD TWIDDLER¿S SYNDROME AND THAT THEY HAD RECEIVED INAPPROPRIATE THERAPY FROM THEIR IMPLANTABLE PULSE GENERATOR (IPG) DEVICE AND LEAD SYSTEM. THE RIGHT ATRIAL (RA) LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338012 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-45

Patients

Seq Age Sex Outcome Treatment
1 00080 YR Hospitalization| R ADDR01 IMPLANTABLE PULSE GENERATOR