FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S MRI SURESCAN

MDR report key: 3861387 · Received June 10, 2014

Report

Report Number
2649622-2014-07298
Event Type
Injury
Date Received
June 10, 2014
Date of Event
March 1, 2013
Report Date
April 15, 2020
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: D314DRM, ICD, IMPLANTED: (B)(6) 2013. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD RISING AND HIGH THRESHOLDS. A CHEST X-RAY SHOWED TENSION ON THE LEAD. THE RV LEAD WAS REPOSITIONED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THE LEAD EXHIBITED UNDERSENSING. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT THE LEAD WAS FRACTURED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
337501 SPRINT QUATTRO SECURE S MRI SURESCAN DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6935M62

Patients

Seq Age Sex Outcome Treatment
1 00061 YR Hospitalization| R 5076-52, LEAD