FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3861329 · Received June 10, 2014

Report

Report Number
2649622-2014-07303
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 15, 2014
Report Date
April 29, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY - THE PARTIAL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED AND NO ANOMALIES WERE FOUND. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION AND ANALYSIS IS PENDING. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE SHOCKS FROM THE LEAD. THE LEAD HAD HIGH IMPEDANCE AND A POSSIBLE FRACTURE. THE LEAD WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340654 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R