VIVA XT
Report
- Report Number
- 3004209178-2014-11245
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- April 5, 2014
- Report Date
- April 6, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY WAS PERFORMED AND NO ANOMALIES WERE FOUND. ANALYST COMMENTED, NO ANOMALIES FOUND.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 5076-52, LEAD, IMPLANTED: (B)(6) 2009; 694765, LEAD, IMPLANTED: (B)(6) 2009. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED DUE TO THE IMPEDANCE ON THE RIGHT VENTRICULAR (RV) LEAD OUT OF RANGE WHICH TRIGGERED A LEAD WARNING. RV AND SVC COILS IMPEDANCES HAVE BEEN ELEVATED AND VARIABLE. INVASIVE ANALYSIS WAS PERFORMED ALONG WITH PROVOCATIVE ISOMETRIC MOVEMENTS. POCKET MANIPULATION REVEALED HIGH AND VARIABLE IMPEDANCES. AN RV COIL CONNECTION ISSUE WAS DISCUSSED. SUBSEQUENT TROUBLESHOOTING AND ANALYSIS REVEALED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) RV HV CONNECTOR WAS LOOSE. THE POCKET WAS OPENED; THE RV HV CONNECTOR WAS REMOVED AND REATTACHED ALONG WITH THE SET SCREW TIGHTENED. CONNECTION AND NORMAL RV LEAD FUNCTION WAS CONFIRMED. THE LEAD AND ICD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 340733 | VIVA XT | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | DTBA1D1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00065 YR | Hospitalization| R | 419678 LEAD |