FDA Adverse Event Injury Summary report: N

ADAPTA DR

MDR report key: 3861168 · Received June 10, 2014

Report

Report Number
3004209178-2014-11254
Event Type
Injury
Date Received
June 10, 2014
Date of Event
March 28, 2014
Report Date
March 28, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NVZ
PMA / PMN Number
P980035
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED NO ANOMALIES WERE FOUND.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 5076-45 LEAD, IMPLANTED: (B)(6) 2013. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS AND WAS EXPLANTED AND REPLACED. DURING THE REPLACEMENT PROCEDURE, WHILE CAUTERY WAS USED THE DEVICE SWITCHED TO UNIPOLAR PACING AND THE PATIENT EXPERIENCED SIGNIFICANT POCKET STIMULATION. THE DEVICE WAS MANUALLY PROGRAMMED TO BIPOLAR AND THE STIMULATION CEASED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COM PLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341693 ADAPTA DR PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ MEDTRONIC MED REL MEDTRONIC PUERTO RICO ADDRL1

Patients

Seq Age Sex Outcome Treatment
1 00050 YR Hospitalization| R 5076-52 LEAD