ADAPTA DR
Report
- Report Number
- 3004209178-2014-11254
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- March 28, 2014
- Report Date
- March 28, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NVZ
- PMA / PMN Number
- P980035
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED NO ANOMALIES WERE FOUND.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 5076-45 LEAD, IMPLANTED: (B)(6) 2013. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS AND WAS EXPLANTED AND REPLACED. DURING THE REPLACEMENT PROCEDURE, WHILE CAUTERY WAS USED THE DEVICE SWITCHED TO UNIPOLAR PACING AND THE PATIENT EXPERIENCED SIGNIFICANT POCKET STIMULATION. THE DEVICE WAS MANUALLY PROGRAMMED TO BIPOLAR AND THE STIMULATION CEASED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COM PLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 341693 | ADAPTA DR | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | ADDRL1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00050 YR | Hospitalization| R | 5076-52 LEAD |