FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3861101 · Received June 10, 2014

Report

Report Number
2649622-2014-07396
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 9, 2014
Report Date
April 9, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 5076-58 IMPLANTABLE PACING LEAD, IMPLANTED: (B)(6) 2013. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED WITH NO ANOMALIES FOUND. THE ANALYST NOTED THAT THE DISTAL LOW VOLTAGE ELECTRODE OF THE LEAD WAS COVERED IN BODY TISSUE/FIBROTIC GROWTH.

Description of Event or Problem · 1

IT WAS REPORTED THAT HIGH THRESHOLD VALUES WERE OBSERVED ON THE RIGHT ATRIAL (RA) LEAD. THE LEAD WAS EXPLANTED AND REPLACED DURING A ROUTINE DEVICE UPGRADE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341565 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 407652

Patients

Seq Age Sex Outcome Treatment
1 00089 YR Required Intervention VEDR01 IMPLANTABLE PULSE GENERATOR