FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3860947
·
Received April 14, 2014
Report
- Report Number
- 1720753-2014-03211
- Event Type
- Malfunction
- Date Received
- April 14, 2014
- Date of Event
- March 24, 2014
- Report Date
- April 14, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE FILAMENT DRIVER PCB AND GENERATOR PCB WERE EVALUATED AND REPLACED. THE HIGH VOLTAGE CABLE WERE EVALUATED AND REGREASED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM EXHIBITED 'PRE-CHARGE VOLTAGE ERRORS' UPON SYSTEM START UP. THIS ERROR IS LIKELY TO PREVENT THE SYSTEM FROM BOOTING TO A USABLE STATE. NO PT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 228453 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |