FDA Adverse Event Injury Summary report: N

PROTECTA XT CRT-D

MDR report key: 3860937 · Received June 10, 2014

Report

Report Number
9614453-2014-01353
Event Type
Injury
Date Received
June 10, 2014
Date of Event
March 24, 2014
Report Date
April 7, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. PRODUCTS: 6935 LEAD IMPLANTED: 2011 (B)(6); 4194 LEAD IMPLANTED: 2011 (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY WAS PERFORMED AND NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE BATTERY HAD A SHORT LIFETIME. THE DEVICE WAS EXPLANTED AND REPLACED. IT WAS FURTHER REPORTED THAT A REVIEW OF THE DEVICE PERFORMANCE DATA INDICATED FIVE ALERTS FOR LOW ATRIAL IMPEDANCE VALUES. THE ATRIAL LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342162 PROTECTA XT CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK IPG MFG SWITZERLAND D354TRG

Patients

Seq Age Sex Outcome Treatment
1 00081 YR Hospitalization| R 4068 LEAD