FDA Adverse Event
Malfunction
Summary report: N
SPRINT QUATTRO SECURE S
MDR report key: 3860924
·
Received June 10, 2014
Report
- Report Number
- 2649622-2014-06821
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- April 18, 2014
- Report Date
- April 18, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT PRIOR TO USE, THE 6935 LEADS APPEARED TO HAVE AIR BUBBLES IN THE OUTER LAYER OF INSULATION. THE INSULATION DEFECT WAS LOCATED ON THE DISTAL PORTION OF THE LEAD AND PROXIMAL TO THE HIGH VOLTAGE COIL ELECTRODE. THE LEADS WERE NEVER IMPLANTED AND RETURNED FOR ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 343065 | SPRINT QUATTRO SECURE S | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 6935M62 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |