FDA Adverse Event Malfunction Summary report: N

ATTAIN PERFORMA

MDR report key: 3860869 · Received June 10, 2014

Report

Report Number
2649622-2014-05274
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
February 12, 2014
Report Date
March 5, 2014
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE U.S. WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THERE WAS BLOOD ON THE DISTAL CONDUCTOR OF THE LEAD AND IT WAS OBSTRUCTED. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT AT THE IMPLANT PROCEDURE THE PHYSICIAN HAD DIFFICULTY INSERTING THE LEFT VENTRICULAR (LV) LEAD INTO THE GUIDEWIRE AND STYLET. THE LV LEAD WAS NOT IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342405 ATTAIN PERFORMA DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 429888

Patients

Seq Age Sex Outcome Treatment
1