FDA Adverse Event Malfunction Summary report: N

XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 3860843 · Received June 10, 2014

Report

Report Number
2024168-2014-03701
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 27, 2014
Report Date
May 15, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED BALLOON RUPTURE WAS UNABLE TO BE CONFIRMED. BASED ON A VISUAL AND FUNCTIONAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A QUERY OF THE ELECTRONIC COMPLAINT DATABASE REVEALED NO OTHER SIMILAR INCIDENTS REPORTED FOR BALLOON RUPTURE FROM THIS LOT. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A 2.75X15MM XIENCE XPEDITION STENT DELIVERY SYSTEM (SDS) WAS ADVANCED, WITHOUT RESISTANCE, TO THE DISTAL LEFT ANTERIOR DESCENDING (LAD) CORONARY ARTERY. DURING INITIAL INFLATION, THE BALLOON RUPTURED AT 10 ATMOSPHERES (ATM). THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT PROCEDURAL DELAY. THERE WAS NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342940 XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 3112841

Patients

Seq Age Sex Outcome Treatment
1 72 YR