FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3860827 · Received April 14, 2014

Report

Report Number
1720753-2014-03236
Event Type
Malfunction
Date Received
April 14, 2014
Date of Event
March 25, 2014
Report Date
April 14, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE HIGH VOLTAGE CABLE CONNECTIONS WERE EVALUATED AND RE-LUBRICATED. A FULL FILAMENT CALIBRATION WAS PERFORMED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM EXHIBITED FILAMENT REGULATOR FAILURE ERRORS DURING A PT PROCEDURE. THIS EVENT MAY HAVE RESULTED IN AN ARO (ACCIDENTAL RADIATION OCCURRENCE). NO PT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
228006 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1