REVEAL XT
Report
- Report Number
- 3004209178-2014-10511
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- April 3, 2014
- Report Date
- April 3, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- DSI
- PMA / PMN Number
- K103764
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
PRODUCT SUMMARY EVENT: THE DEVICE WAS NOT RETURNED FOR ANALYSIS; HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED OVERSENSING ON DUE TO EMI (ELECTROMAGNETIC INTERFERENCE) AND EPISODES WITH APPARENT NOISE ON THE ELECTROGRAM.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) WAS DETECTING NOISE IN THE SENSING CHANNEL AND WAS UNABLE TO SENSE ANY R-WAVES. THE PROGRAMMED ICM SENSITIVITY WAS ADJUSTED TO NO AVAIL. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 338806 | REVEAL XT | DETECTOR AND ALARM, ARRHYTHMIA | DSI | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | 9529 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00073 YR | Required Intervention |