FDA Adverse Event
Malfunction
Summary report: N
PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
MDR report key: 3860806
·
Received June 10, 2014
Report
- Report Number
- 2183613-2014-00758
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- April 8, 2014
- Report Date
- April 14, 2014
- Manufacturer
- MEDTRONIC MILACA, INC.
- Product Code
- LWP
- PMA / PMN Number
- P820003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT, THE VENTRICULAR OUTPUT CONNECTOR WAS BROKEN. IT WAS ALSO NOTED THAT ONE SIDE BAIL COVER WAS BROKEN, THE BATTERY CONTACTS WERE COMPRESSED, THE RING WAS BENT, AND THE KEYBOARD PAD WAS DAMAGED AND SCRATCHED.
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Additional Manufacturer Narrative · 1
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR HAD A BROKEN VENTRICULAR OUTPUT CONNECTOR. THE GENERATOR WAS RETURNED FOR SERVICE. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 338804 | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | MEDTRONIC MILACA, INC. | 5388 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |