FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 3860753
·
Received June 10, 2014
Report
- Report Number
- 2649622-2014-05314
- Event Type
- Injury
- Date Received
- June 10, 2014
- Report Date
- March 6, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: ADDR01 IPG IMPLANTED: (B)(6) 2001.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS WHILE THE PATIENT WAS SITTING UPRIGHT. IN THE SUPINE POSITION, THRESHOLDS AND IMPEDANCES WERE NORMAL; HOWEVER, THE THRESHOLDS WERE HIGH WHEN THE PATIENT SAT UPRIGHT. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 337288 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-45 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00088 YR | Hospitalization| R | 4092 LEAD |