FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

MDR report key: 3860737 · Received June 10, 2014

Report

Report Number
2182208-2014-01449
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 24, 2014
Report Date
April 24, 2014
Manufacturer
MEDTRONIC, INC.
Product Code
LWP
PMA / PMN Number
P820003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; LCD (LIQUID CRYSTAL DISPLAY) MISSING SEGMENT. ANALYSIS ALSO FOUND THE BATTERY CONTACTS ARE COMPRESSED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BOTTOM DISPLAY OF THE EXTERNAL PULSE GENERATOR (EPG) WOULD NOT LIGHT UP COMPLETELY. IT WAS ALSO REPORTED THE BOTTOM SCREEN IS DEFECTIVE, MENU CANNOT BE READ. THE EPG WAS RETURNED FOR REPAIR. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
337186 PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MEDTRONIC, INC. 5388

Patients

Seq Age Sex Outcome Treatment
1