INSYNC III
Report
- Report Number
- 9614453-2014-01160
- Event Type
- Injury
- Date Received
- June 10, 2014
- Report Date
- March 4, 2014
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- LWP
- PMA / PMN Number
- P010015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCTS: 5076-52 IMPLANTABLE PACING LEAD 2008 (B)(6); 5076-58 IMPLANTABLE PACING LEAD 2008 (B)(6). (B)(4).
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE ACTUAL LONGEVITY WAS LESS THAN 80% OF THE 99.9% LONGEVITY LIMIT. THE DEVICE WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) AND HAS POSSIBLE PREMATURE BATTERY DEPLETION. THE LEFT VENTRICULAR (LV) LEAD STOPPED CAPTURING ONE YEAR AFTER IMPLANT. THE DEVICE WAS EXPLANTED AND THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 340012 | INSYNC III | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | IPG MFG SWITZERLAND | 8042B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00072 YR | Hospitalization| R | 419588 IMPLANTABLE PACING LEAD |