FDA Adverse Event Injury Summary report: N

INSYNC III

MDR report key: 3860720 · Received June 10, 2014

Report

Report Number
9614453-2014-01160
Event Type
Injury
Date Received
June 10, 2014
Report Date
March 4, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
LWP
PMA / PMN Number
P010015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCTS: 5076-52 IMPLANTABLE PACING LEAD 2008 (B)(6); 5076-58 IMPLANTABLE PACING LEAD 2008 (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE ACTUAL LONGEVITY WAS LESS THAN 80% OF THE 99.9% LONGEVITY LIMIT. THE DEVICE WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) AND HAS POSSIBLE PREMATURE BATTERY DEPLETION. THE LEFT VENTRICULAR (LV) LEAD STOPPED CAPTURING ONE YEAR AFTER IMPLANT. THE DEVICE WAS EXPLANTED AND THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340012 INSYNC III PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP IPG MFG SWITZERLAND 8042B

Patients

Seq Age Sex Outcome Treatment
1 00072 YR Hospitalization| R 419588 IMPLANTABLE PACING LEAD