FDA Adverse Event
Malfunction
Summary report: N
GEMSTR 7 THPY ALT ST
MDR report key: 3860690
·
Received April 15, 2014
Report
- Report Number
- 9615050-2014-02693
- Event Type
- Malfunction
- Date Received
- April 15, 2014
- Date of Event
- March 18, 2014
- Report Date
- March 19, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K060806
- Removal / Correction Number
- Z-1159-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TESTING AND INVESTIGATION FOUND THE DEVICE ALARMED FOR SERVICE ALARM CODE 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY). THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE RPTR UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE ALARMED WITH AN 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE ALARM CODE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 230557 | GEMSTR 7 THPY ALT ST | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |