FDA Adverse Event Injury Summary report: N

VIVA XT

MDR report key: 3860648 · Received June 10, 2014

Report

Report Number
3004209178-2014-10534
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 21, 2014
Report Date
April 21, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY #(B)(4): THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE INDICATED A DAMAGED GROMMET. ANALYSIS OF THE DEVICE MEMORY INDICATED OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS/SIC (SENSING INTEGRITY COUNTER).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 429688, PACING LEAD, IMPLANTED: (B)(6) 2014; 5076-45, PACING LEAD, IMPLANTED: (B)(6) 2012. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT POST IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM IMPLANT AND POST DEFIBRILLATION TESTING, 3 BRIEF MECHANICAL TYPE NOISE OVERSENSING VENTRICULAR EPISODES WERE OBSERVED. THE POCKET WAS RE-OPENED AND THE CONNECTIONS WERE VERIFIED. THE ICD WAS PLACED BACK IN THE POCKET AND ANOTHER EPISODE OF VENTRICULAR OVERSENSING WAS THEN OBSERVED. THE ICD AND RIGHT VENTRICULAR (RV) LEAD WERE REMOVED AND REPLACED. THE PATIENT DEVELOPED A HEMATOMA THAT WAS ATTRIBUTED TO GOING IN THE SECOND TIME TO REVISE THE SYSTEM. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
339638 VIVA XT DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC PUERTO RICO OPERATIONS CO. DTBA1D4

Patients

Seq Age Sex Outcome Treatment
1 00072 YR Hospitalization| R 6947M62 TACHY LEAD