FDA Adverse Event
Injury
Summary report: N
SPRINT QUATTRO SECURE S
MDR report key: 3860625
·
Received June 10, 2014
Report
- Report Number
- 2649622-2014-05363
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- March 7, 2014
- Report Date
- March 7, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 5076 LEAD, (B)(6) 2007. (B)(4).
Additional Manufacturer Narrative · 1
EVENT SUMMARY - THE PARTIAL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. ANALYSIS REVEALED THAT THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD AN ABRUPT INCREASE IN THRESHOLD. THE LEAD ALSO HAD A POSSIBLE FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 340364 | SPRINT QUATTRO SECURE S | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 693565 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00044 YR | Hospitalization| R | D314DRG ICD |