FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 3860625 · Received June 10, 2014

Report

Report Number
2649622-2014-05363
Event Type
Injury
Date Received
June 10, 2014
Date of Event
March 7, 2014
Report Date
March 7, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 5076 LEAD, (B)(6) 2007. (B)(4).

Additional Manufacturer Narrative · 1

EVENT SUMMARY - THE PARTIAL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. ANALYSIS REVEALED THAT THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD AN ABRUPT INCREASE IN THRESHOLD. THE LEAD ALSO HAD A POSSIBLE FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340364 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 693565

Patients

Seq Age Sex Outcome Treatment
1 00044 YR Hospitalization| R D314DRG ICD