FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 3860589 · Received June 10, 2014

Report

Report Number
2649622-2014-05375
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
March 14, 2014
Report Date
March 14, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. THE HELIX OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING /STRETCHING/OVERSTRESS AND WAS EXTRINSICALLY BENT. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE U.S. WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THE LEAD HELIX EXTENDED ON ITS OWN DURING THE PROCEDURE, AND IT WAS NOT POSSIBLE TO RETRACT THE HELIX. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341755 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6935M62

Patients

Seq Age Sex Outcome Treatment
1