FDA Adverse Event Malfunction Summary report: N

ADAPTA DR

MDR report key: 3860544 · Received June 10, 2014

Report

Report Number
9614453-2014-01168
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 22, 2014
Report Date
April 22, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NVZ
PMA / PMN Number
P980035
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT FOLLOWING A DEVICE CHECK, THE PATIENT FELT PAIN CAUSED BY TWITCHING OF THE POCKET SITE. THE EXACT CAUSE WAS EVENTUALLY DETERMINED TO BE THAT THE PROGRAMMER EMERGENCY BUTTON HAD BEEN PUSHED INADVERTENTLY DURING THE DEVICE CHECK. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341654 ADAPTA DR PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ IPG MFG SWITZERLAND ADDRS1

Patients

Seq Age Sex Outcome Treatment
1 00090 YR