FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3860510 · Received April 15, 2014

Report

Report Number
1720753-2014-03270
Event Type
Malfunction
Date Received
April 15, 2014
Date of Event
March 23, 2014
Report Date
April 15, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE F4 FUSE ON THE POWER RELAY PCB WAS EVALUATED AND IDENTIFIED AS REQUIRING REPLACEMENT. NO CONCLUSION CAN BE DRAWN AS CONCLUSIVE REPAIR INFO IS UNAVAILABLE AT THIS TIME AND NO ADDITIONAL SERVICE INFO WAS PROVIDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM EXHIBITED 'PRE-CHARGE ERRORS' UPON SYSTEM START UP. THIS ERROR IS LIKELY TO PREVENT THE SYSTEM FROM BOOTING TO A USABLE STATE. NO PT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
230620 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1