FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3860501
·
Received April 15, 2014
Report
- Report Number
- 1720753-2014-03262
- Event Type
- Malfunction
- Date Received
- April 15, 2014
- Date of Event
- March 27, 2014
- Report Date
- April 15, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE X-RAY SWITCH WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM PRODUCED UNCOMMANDED X-RAYS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.