SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2014-05422
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- March 8, 2014
- Report Date
- March 10, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Removal / Correction Number
- Z-0475-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY - THE FULL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. THE PROXIMAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION AND ANALYSIS IS PENDING. CONCOMITANT MEDICAL PRODUCTS: D274TRK ICD, (B)(6) 2011; 4076 LEAD, (B)(6) 2013. (B)(4).
IT WAS REPORTED THAT THE PATIENT RECEIVED FOUR INAPPROPRIATE SHOCKS DUE TO OVERSENSING ON THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD ALSO HAS A POSSIBLE FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 342026 | SPRINT QUATTRO SECURE | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 694765 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00063 YR | Hospitalization| L| R | 4296 LEAD |