FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3860478 · Received June 10, 2014

Report

Report Number
2649622-2014-05422
Event Type
Injury
Date Received
June 10, 2014
Date of Event
March 8, 2014
Report Date
March 10, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY - THE FULL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. THE PROXIMAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION AND ANALYSIS IS PENDING. CONCOMITANT MEDICAL PRODUCTS: D274TRK ICD, (B)(6) 2011; 4076 LEAD, (B)(6) 2013. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED FOUR INAPPROPRIATE SHOCKS DUE TO OVERSENSING ON THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD ALSO HAS A POSSIBLE FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342026 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00063 YR Hospitalization| L| R 4296 LEAD