FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 3860464 · Received April 15, 2014

Report

Report Number
1720753-2014-03290
Event Type
Malfunction
Date Received
April 15, 2014
Date of Event
March 24, 2014
Report Date
April 15, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE INTERCONNECT CABLE WAS EVALUATED AND REPLACED. THE HV TANK PCB CONNECTIONS WERE ALSO RESEATED AS PART OF THE SERVICE EVENT. THE SYSTEM WAS TESTED AND FOUND TO WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY FAILED TO ALLOW FLUOROSCOPIC EXPOSURES AND EXHIBITED INTERMITTENT FUNCTIONALITY. NO PT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
230726 8800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 8800

Patients

Seq Age Sex Outcome Treatment