FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY PLUS

MDR report key: 3860283 · Received June 10, 2014

Report

Report Number
2649622-2014-05459
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
March 12, 2014
Report Date
March 12, 2014
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. THE ANALYST COMMENTED, ELECTRICAL RESISTANCE AND CROSS CONTINUITY TEST RESULTS WERE WITHIN SPECIFICATIONS. NO ANOMALIES WERE FOUND. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE THE PATIENT EXPERIENCED PHRENIC NERVE/DIAPHRAGMATIC MUSCLE STIMULATION AND THE LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH THRESHOLD. PER MEDICAL JUDGEMENT THE LV LEAD WAS REPLACED WITH A DIFFERENT LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
343350 ATTAIN ABILITY PLUS DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 429688

Patients

Seq Age Sex Outcome Treatment
1 00068 YR