FDA Adverse Event
Malfunction
Summary report: N
ATTAIN ABILITY PLUS
MDR report key: 3860283
·
Received June 10, 2014
Report
- Report Number
- 2649622-2014-05459
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- March 12, 2014
- Report Date
- March 12, 2014
- Manufacturer
- MPRI
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. THE ANALYST COMMENTED, ELECTRICAL RESISTANCE AND CROSS CONTINUITY TEST RESULTS WERE WITHIN SPECIFICATIONS. NO ANOMALIES WERE FOUND. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE THE PATIENT EXPERIENCED PHRENIC NERVE/DIAPHRAGMATIC MUSCLE STIMULATION AND THE LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH THRESHOLD. PER MEDICAL JUDGEMENT THE LV LEAD WAS REPLACED WITH A DIFFERENT LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 343350 | ATTAIN ABILITY PLUS | DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE | OJX | MPRI | 429688 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00068 YR |