FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 3860267 · Received June 10, 2014

Report

Report Number
2649622-2014-05513
Event Type
Injury
Date Received
June 10, 2014
Date of Event
March 14, 2014
Report Date
March 21, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. THE OUTER AND INNER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A CUT. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. THE ANALYST COMMENTED, DUE TO THE LENGTH OF IMPLANT AND THE AMOUNT OF BLOOD INGRESS IT IS LIKELY THAT THE INNER/OUTER INSULATION EXTRINSIC BREACH/CUT OCCURRED DURING THE IMPLANT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS UNDERSENSING R WAVES. THE PHYSICIAN DECIDED TO EXPLANT AND REPLACE THE RV LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
343253 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6935M55

Patients

Seq Age Sex Outcome Treatment
1 00070 YR Hospitalization| R DVBB1D4, ICD